When does AI software become a medical device?

A medical device manufacturer, hospital group or health software vendor that plugs Claude or another model into its processes eventually asks whether the tool falls under Regulation (EU) 2017/745 on medical devices, the MDR, and therefore under the AI Act. The answer turns on what the company says the tool is for.

AI software is a medical device when its manufacturer intends it for one of the medical purposes listed in Article 2(1) of the MDR, such as diagnosis, prevention, monitoring, prediction, prognosis, treatment or alleviation of disease. Qualification follows the intended purpose declared by the manufacturer, and Rule 11 sets the class.

Intended purpose is set by the manufacturer

The MDR defines intended purpose as the use for which a device is intended according to the manufacturer's label, instructions for use, promotional or sales materials or statements, and clinical evaluation (Article 2(12)). A sales brochure promising to "support diagnosis" therefore weighs on qualification as much as the instructions for use. MDCG 2019-11, revised in June 2025 by the Medical Device Coordination Group, makes the point: an ambiguous intended purpose can lead to misclassification, and every medical claim needs an appropriate level of clinical evidence behind it.

The same guidance excludes software intended only for non-medical purposes and names examples: invoicing, staff planning, e-mailing, word processing, back-up, wellness and fitness apps. It also states that the risk of harm linked to using software in a healthcare setting is not, in itself, a criterion for qualifying it as a medical device.

Simple search versus processing for a medical purpose

Retrieving records by matching metadata is "simple search", which does not make software a medical device. Software that processes, analyses, interprets, creates or modifies medical information may qualify as soon as that operation serves a medical purpose, and section 3.1 of the guidance specifically mentions natural language processing search among the functions concerned. An agent that summarises a patient record to inform a clinical decision may therefore fall under the MDR. The same agent preparing the invoice for that hospital stay stays outside.

How does MDR Rule 11 classify AI software?

Once qualified, software is classified under Annex VIII of the MDR, where Rule 11 deals specifically with software. It reads as a cascade:

  1. software intended to provide information used to take decisions with diagnostic or therapeutic purposes is class IIa;
  2. it moves to class IIb if those decisions may cause a serious deterioration of health or a surgical intervention, and to class III if they may cause death or an irreversible deterioration;
  3. software intended to monitor physiological processes is class IIa, or class IIb when it monitors vital parameters whose variations could result in immediate danger to the patient;
  4. all other software is class I.

Class IIa is where decision support usually lands

MDCG 2019-11 notes that the wording "intended to provide information which is used to take decisions with diagnosis or therapeutic purposes" describes, in very general terms, the mode of action of all medical device software. In our reading, a clinical decision-support agent almost always starts in class IIa, and class I remains the exception for software with a medical purpose.

Why the class also matters for the AI Act

The MDR reserves self-certification for class I, except for sterile devices, devices with a measuring function and reusable surgical instruments (Article 52(7)). From class IIa upwards, a notified body takes part in the conformity assessment (Article 52(6)), and that involvement is exactly what the AI Act uses to classify the software as high-risk.

When is a medical device a high-risk AI system?

Article 6(1) of the AI Act (Regulation (EU) 2024/1689) classifies an AI system as high-risk when it is itself a product covered by Annex I, or a safety component of such a product, and that product requires a third-party conformity assessment. The MDR and Regulation (EU) 2017/746 on in vitro diagnostic medical devices sit in Section A of Annex I, as points 11 and 12.

The notified body as the trigger

The joint guidance MDCG 2025-6, adopted in June 2025 with the European Artificial Intelligence Board, sets out the result in a table: self-certified class I AI software is not high-risk, whereas class IIa, IIb and III software is. The MDR class of the device determines its status under Article 6(1). The guidance predates the Omnibus, so its dates need re-reading.

In-house tools built by health institutions

A hospital that designs and uses its own software under the conditions of Article 5(5) of the MDR does not go through a notified body. MDCG 2025-6 concludes that such an in-house tool is not high-risk, while recalling that the rest of the AI Act, prohibited practices included, still applies.

What did the Digital Omnibus change for medical devices?

Regulation (EU) 2026/1744, the Digital Omnibus on AI, published in the Official Journal on 24 July 2026 and in force since 27 July, keeps the MDR and the IVDR in Section A of Annex I. It changes the timetable and the way the two regimes fit together.

2 August 2028 for Annex I systems

Sections 1 to 3 of Chapter III, which carry the obligations for high-risk systems, apply from 2 August 2028 to systems classified under Article 6(1) and Annex I (amended Article 113), against 2 December 2027 for Annex III systems. Rewritten Article 111(2) adds that a system placed on the market before the date that applies to it only falls under the Regulation if its design changes significantly afterwards.

Article 2(13): an equivalence still to be defined

The Omnibus adds a paragraph 13 to Article 2. For high-risk systems under Article 6(1), the application of Articles 9 to 15 and 17 to 25 may be limited where the Section A legislation provides an equivalent or higher level of protection of health, safety or fundamental rights, provided the limitation does not reduce the overall level of protection. The Commission must list the systems and requirements concerned in delegated acts by 2 August 2027. As of 7 October 2026, those acts have not been adopted. We recommend documenting now, requirement by requirement, what the MDR file already covers.

One conformity assessment, run under the MDR procedure

Rewritten Article 43(3) confirms that the manufacturer follows the MDR procedure and that the requirements of Chapter III, Section 2 of the AI Act form part of that assessment. Notified bodies designated under the MDR must apply for designation under the AI Act by 28 January 2028. Unchanged since 2024, Article 74(3) hands oversight to the authority designated under the MDR, and Article 73(10) limits AI Act serious incident reporting for these devices to fundamental rights breaches. In France, the scheme presented by the government in September 2025 assigns these devices to ANSM together with the DGCCRF, and the designating law has not been adopted as of 7 October 2026.

The December 2025 proposal, still under negotiation

On 16 December 2025 the Commission proposed a targeted revision of the MDR and the IVDR (COM(2025) 1023). Its Article 4 moves both regulations to Section B of Annex I to the AI Act: Article 6(1) would still classify this software as high-risk, but its requirements would come from acts adopted under the MDR. The 7 May 2026 political agreement on the Omnibus kept the MDR in Section A, with Article 2(13) as the compromise, and the Section B route remains under discussion in this separate negotiation (Council progress report of 5 June 2026, point 19). As of 10 November 2026 the text has not been adopted. Announced next steps: a vote in the European Parliament's Committee on Public Health, expected between December 2026 and early 2027 (no date is listed in the Legislative Observatory as of that date), a Council general approach targeted for the end of 2026 and a Parliament position expected in early 2027. As long as this revision is not adopted and applicable, class IIa and higher AI software falls under all high-risk requirements, applicable from 2 August 2028, subject to the Article 2(13) acts.

Which AI agents stay outside the MDR?

An agent with no medical purpose falls under general rules: AI Act Article 50, applicable since 2 August 2026, if it talks to people, Article 4 on AI literacy, the GDPR as soon as it touches personal data, and internal governance. Health data brings its own requirements, covered on our AI agents for healthcare and life sciences page.

Technical documentation and regulatory files

An agent can compare two versions of the instructions for use, check that each general safety and performance requirement points to evidence, or prepare the traceability matrix the notified body will review. It works on the file without taking part in how the device functions. An error by the agent can still end up in a file submitted to the notified body, and the person responsible for regulatory compliance under Article 15 of the MDR keeps responsibility for what goes out.

Quality, complaints and vigilance

Within the quality management system, an agent can group logged complaints by product family, or match a nonconformity with past corrective actions. Vigilance is the sensitive part. Deciding that an event is a serious incident to report to the competent authority remains a human decision, taken by the designated person, to whom the agent hands every ambiguous signal without classifying it.

Monitoring, back office and support functions

An automated regulatory monitoring agent flags publications that affect a product range and drafts an impact note that the regulatory affairs team approves. Invoicing and staff planning, for their part, appear by name in the MDCG 2019-11 list of non-medical software.

Where is the line for an agent close to the patient?

When the agent's output changes nature

An agent that drafts a discharge letter from clinical notes, reviewed and signed by the physician, is in principle doing writing work on health data. Once it proposes a differential diagnosis, flags a patient at risk of deterioration or suggests a dosage, it provides information used for diagnostic or therapeutic decisions and falls under Rule 11. The shift can come from a feature added mid-project at users' request, and that is the moment to run qualification again.

A general-purpose model steered towards a medical purpose

A general-purpose AI model only becomes a medical device once someone gives it a medical intended purpose. Whoever places a diagnostic support tool built on that model on the market under its own name becomes its manufacturer within the meaning of Article 2(30) of the MDR. On the AI Act side, Article 25(1)(c) turns a deployer into a provider when it changes the intended purpose of an AI system, general-purpose systems included, in a way that makes it high-risk. The original provider must then cooperate unless it clearly ruled out that transformation (Article 25(2)), and the Omnibus spelled out what cooperation covers: documentation, known limitations, technical access. The company that sets the medical intended purpose carries the manufacturer's file, whatever model sits underneath.

How do you govern these agents with LOOP™?

LOOP™, Koneetiv's governance method for AI agents, assigns every action an agent can take to one of 4 trust zones before go-live. At a medical device manufacturer, the regulatory line maps directly onto that classification:

This block turns Rule 11 into a technical safeguard: qualification is decided by the regulatory committee, and no conversation with the agent can change it. Every agent is entered in the LOOP™ living register, with the data it accesses, its versioned instructions and its scope. We recommend attaching its MDR qualification conclusion and rationale. The logic is explained in our article on sorting AI agents into trust zones, and the method on the LOOP™ AI agent governance page.

Where should a manufacturer or hospital group start?

Before opening an agent project, we suggest working in this order:

  1. List current and planned AI uses, consumer tools included.
  2. Write each tool's intended purpose in one sentence and test it against Article 2 of the MDR and the MDCG 2019-11 examples.
  3. For qualified software, apply Rule 11 and put the 2 August 2028 AI Act date into the quality plan.
  4. For the rest, check Article 50, AI literacy and the GDPR, then sort each action into a trust zone.
  5. Review sales materials, instructions for use and on-screen text, where a medical promise can change the intended purpose.
  6. Track the Article 2(13) delegated acts and the MDR revision.

The wider timetable is covered in our AI Act compliance guide and in the article on the August 2026 deadline for leadership teams. Alongside an ISO 13485 quality system, the ISO/IEC 42001 AI management standard can hold the AI use policy and the periodic review of agents.

Koneetiv is a French Claude-only consultancy and an official Anthropic partner working with enterprises and regulated industries. We scope these projects with quality and regulatory affairs teams, whatever the size of the company. To measure the maturity of your agent governance first, the AI maturity assessment places you on the 6 axes of the Koneetiv framework (2026 edition).